Sensor type
|
F0,5 (HD)
|
N0,2 (HD)
|
F1.5, N0.7, FN1.5 (HD)
|
F1.5-90, N0.7-90, FN1.5-90
|
F2 HD
|
F5, N2.5, FN5 (HD)
|
N7
|
F15
|
F35
|
Properties
|
F
|
N
|
F
|
N
|
F
|
N
|
F
|
F
|
N
|
N
|
F
|
F
|
Measuring range
|
0 ... 0,5 m
|
0 ... 0,2 m
|
0…1,5 mm
|
0…0,7 mm
|
0…1,5 mm
|
0…0,7 mm
|
0…2 mm
|
0…5 mm
|
0…2,5 mm
|
0…7 mm
|
0…15 mm
|
0…35 mm
|
Field of application
|
Small samples, thin coatings, use with measuring stand
|
Small samples, thin coatings, use with measuring stand
|
Tubes and pipes or objects difficult to access
|
Rough surfaces
|
Standard sensors for a wide field of applications
|
Thick coatings, wall thickness
|
Thick coatings, wall thickness
|
Thick coatings, wall thickness
|
Measuring principle
|
Magnetic induction
|
Eddy currents
|
Magnetic induction
|
Eddy currents
|
Magnetic induction
|
Eddy currents
|
Magnetic induction
|
Magnetic induction
|
Eddy currents
|
Eddy currents
|
Magnetic induction
|
Magnetic induction
|
Signal processing
|
Sensor-integrated 32 bits signal processing (SIDSP®)
|
Accuracy
|
± (1 μm + 0,75% of reading)
|
± (1,5 μm + 0,75% of reading)
|
± (5 μm + 0,75% of reading)
|
± (5 μm + 0,75% of reading)
|
± (20 μm + 0,75% of reading)
|
Repeatability
|
± (0,5 μm + 0,5% of reading)
|
± (0,8 μm + 0,5% of reading)
|
± (2,5 μm + 0,5% of reading)
|
± (2,5 μm + 0,5% of reading)
|
± (10 μm + 0,5% of reading)
|
Low range resolution
|
0,02 μm
|
0,05 μm
|
0,1 μm
|
1 μm
|
1 μm
|
5 μm
|
Minimum curvature radius convex
|
1,0 mm
|
1,0 mm
|
-
|
1,5 mm
|
15 mm
|
5 mm
|
50 mm
|
Minimum curvature radius concave
|
7,5 mm
|
7,5 mm
|
5 mm
|
10 mm
|
25 mm
|
25 mm
|
50 mm
|
Minimum measuring area
(without / with stand)
|
Ø14 / Ø5 mm
|
Ø14 / Ø5 mm
|
Ø14 mm
|
Ø14 mm
|
Ø14 / Ø10 mm
|
Ø23 / Ø20 mm
|
Ø25 / Ø25 mm
|
Ø100 mm
|
Minimum substrate thickness
|
0,3 mm
|
40 μm
|
0,3 mm
|
40 μm
|
0,3 mm
|
40 μm
|
0,5 mm
|
0,5 mm
|
40 μm
|
40 μm
|
1 mm
|
1,5 mm
|
Sortie câble
|
0/90°
|
0/90°
|
0/90°
|
0/90°
|
0°
|
0°
|
0°
|
0/90°
|
0/90°
|
0/90°
|
0/90°
|
0/90°
|
Measuring rate in continuous mode
|
20 readings per second
|
Max. measuring rate in single readings mode
|
70 readings per minute
|