content="14/09/2022″/> Force transducer DMS 730, ± 0,1 %, 50 KN ~ 2 MN

Force transducer DMS 730, ± 0,1 %, 50 KN ~ 2 MN

Force transducer DMS 730, ± 0,1 %, 50 KN ~ 2 MN
Article code730
- Fmax: 50 kN to 2 MN, higher forces on request
- suitable for compression force  
- can be used for static and dynamic force
- parallel connection of several transducers possible (Version K)
- protection class IP 65.
- housing nickel plated, optional in stainless steel or aluminium.
- customized versions available

Characteristics
- combined error:
 K = ± 0,1 % Fmax
 N = ± 0,25 % Fmax
- temperature effect per 10 K
 a) of zero:
    K = ± 0,05 % Fmax
    N = ± 0,1 % Fmax
 b) of sensitivity:
    K = ± 0,1 % Fmax
    N = ± 0,2 % Fmax
- input resistance: 350 ± 3 Ω
- output resistance: 350 ± 3 Ω
- insulation resistance: > 4 GΩ
- zero balance: 0,5 %
- recommended supply voltage: 10 V
- nominal range of supply voltage: 2...12 V
- maximum supply voltage: 15 V
- service load: 120 % Fmax
- max permissible load: 150 % Fmax
- breaking load: > 300 % Fmax
- maximum side load: 110 % Fmax
- max permissible dynamic load: 70 % Fmax
- displacement at nominal load: 0,2 mm
- reference temperature: +23°C
- nominal temperature range: -10 ... +50°C
- service temperature range: -20 ... +70°C
- storage temperature: -20 ... +80°C
- protection class (DIN 40050): IP 65

For more information, click here